A Kibble balance (also formerly known as a watt balance) is an electromechanical measuring instrument that measures the weight of a test object very precisely by the electric current and voltage needed to produce a compensating force. It is a Metrology instrument that can realize the definition of the kilogram unit of mass based on fundamental constants.
It was originally known as a watt balance because the weight of the test mass is proportional to the product of current and voltage, which is measured in . In June 2016, two months after the death of its inventor, Bryan Kibble, metrologists of the Consultative Committee for Units of the International Committee for Weights and Measures agreed to rename the device in his honor. Consultative Committee for Units (CCU), Report of the 22nd meeting (15-16 June 2016), pp. 32-32, 35
Prior to 2019, the definition of the kilogram was based on a physical object known as the International Prototype of the Kilogram (IPK). After considering alternatives, in 2013 the General Conference on Weights and Measures (CGPM) agreed on accuracy criteria for replacing this definition with one based on the use of a Kibble balance. After these criteria had been achieved, the CGPM voted unanimously on November 16, 2018, to change the definition of the kilogram and several other units, effective May 20, 2019, to coincide with World Metrology Day. Decision CIPM/105-13 (October 2016) There is also a method called the joule balance. All methods that use the fixed numerical value of the Planck constant are sometimes called the Planck balance.
The main weakness of the ampere balance method is that the result depends on the accuracy with which the dimensions of the coils are measured. The Kibble balance uses an extra calibration step to cancel the effect of the geometry of the coils, removing the main source of uncertainty. This extra step involves moving the force coil through a known magnetic flux at a known speed. This was possible by setting of the conventional values of the von Klitzing constant and Josephson constant, which are used throughout the world for voltage and resistance calibration. Using these principles, in 1990 Bryan Kibble and Ian Robinson invented the Kibble Mark II balance, which uses a circular coil and operates in vacuum conditions . Bryan Kibble worked with Ian Robinson and Janet Belliss to build this Mark Two version of the balance. This design allowed for measurements accurate enough for use in the redefinition of the SI unit of mass: the kilogram.
The Kibble balance originating from the National Physical Laboratory was transferred to the National Research Council of Canada (NRC) in 2009, where scientists from the two labs continued to refine the instrument. In 2014, NRC researchers published the most accurate measurement of the Planck constant at that time, with a relative uncertainty of 1.8. A final paper by NRC researchers was published in May 2017, presenting a measurement of the Planck constant with an uncertainty of only 9.1 parts per billion, the measurement with the least uncertainty to that date. Other Kibble balance experiments are conducted in the US National Institute of Standards and Technology (NIST), the Swiss Federal Office of Metrology (METAS) in Berne, the International Bureau of Weights and Measures (BIPM) near Paris and Laboratoire national de métrologie et d'essais (LNE) in Trappes, France.
While "weighing", the system measures , by controlling the current in the coil to keep the electromagnetic force on the coil balanced with the force of gravity. Coil position and velocity measurement circuitry uses an interferometer together with a precision clock input to determine the velocity and control the current needed to maintain it. The required current is measured, using an ammeter comprising a Josephson junction voltage standard and an integrating voltmeter.
While "moving", the system measures and , by ceasing to provide current to the coil. This allows the counterbalance to pull the coil (and mass) upward through the magnetic field, which causes a voltage difference across the coil. The velocity measurement circuitry measures the speed of movement of the coil. This voltage is measured, using the same voltage standard and integrating voltmeter.
A typical Kibble balance measures , , and , but does not measure the local gravitational acceleration , because does not vary rapidly with time. Instead, is measured in the same laboratory using a highly accurate and precise gravimeter. In addition, the balance depends on a highly accurate and precise frequency reference such as an atomic clock to compute voltage and current. Thus, the precision and accuracy of the mass measurement depends on the Kibble balance, the gravimeter, and the clock.
Like the early atomic clocks, the early Kibble balances were one-of-a-kind experimental devices and were large, expensive, and delicate. As of 2019, work is underway to produce standardized devices at prices that permit use in any metrology laboratory that requires high-precision measurement of mass.
As well as large Kibble balances, microfabricated or MEMS watt balances (now called Kibble balances) have been demonstrated since around 2003. These are fabricated on single silicon dies similar to those used in microelectronics and accelerometers, and are capable of measuring small forces in the nanonewton to micronewton range traceably to the SI-defined physical constants via electrical and optical measurements. Due to their small scale, MEMS Kibble balances typically use electrostatic rather than the inductive forces used in larger instruments. Lateral and torsional variants have also been demonstrated, with the main application (as of 2019) being in the calibration of the atomic force microscope. Accurate measurements by several teams will enable their results to be averaged and so reduce the experimental error.
The principle of the electronic kilogram relies on the value of the Planck constant, which is as of 2019 an exact value. This is similar to the metre being defined by the speed of light. With the constant defined exactly, the Kibble balance is not an instrument to measure the Planck constant, but is instead an instrument to measure mass:
|
|